Lullaby Resus Prime

Lullaby Resus Prime

Effective suction. Controlled resuscitation. One solution.

Contact Us
Category:
Description
The Lullaby Resus Prime supports
• Oral/nasal pharyngeal suctioning as well as tracheal suction to patients.
• Monitored delivery of positive pressure through a manual resuscitation bag.†
• Clinical accuracy in pressure delivery.
Technical Specifications
Environmental specifications • Operating temperature 18° to 41°C
• Operating humidity 0 to 90% Non-Condensing RH
• Storage temperature –25° to 60°C
• Storage humidity 0 to 95% Non-Condensing RH
• Storage pressure 50 to 106 kPa
Physical characteristics • Depth: 140 mm
• Height: 222 mm
• Width: 195 mm
• Weight Approx. 3 kgs
System input requirements • Input gas pressure (wall): 40-80 psi (275-550 kPa) Gas supply minimum flow
rating:70L/min High-pressure cylinders: 3000 psi max.
System characteristics • Flow range: 0-15 LPM
• Flow accuracy: ±2 LPM
Venturi Based Suction • Vacuum Range: 0 to 150±10 mmHg
• Manometer Range: 0 to 200 mmHg
• Manometer accuracy ±5% of full scale
Vacuum Suction accuracy • Set vacuum value** will have accuracy of ±10 mmHg across 40 to 80 psi range.
User control settingsy • O2 flowmeter 0-15 lpm set range
Recommended patient body weight
range
• Up to 7 kg
Approximate operating time
during suctioning (10L cylinder)
• >70 minutes
(at 100 mmHg and with oxygen flow meter set to OFF)
Standards • This device is ISO 10079-3 qualified.