Giraffe Warmers

Giraffe Warmers

The Giraffe Shuttle is a transportable power source that is an accessory to Giraffe and Panda1 beds. It connects to the bed and provides electrical power to the bed and other auxiliary equipment, as well as provisions for medical gas management, required for patient care during intra-hospital transport.2 The Giraffe Shuttle facilitates the mobility of these beds within a hospital building.

Contact Us
Category:
Description
Giraffe Warmer • Height: (adjustable) 193-218 cm (76-86 in)
• Depth: 119.4 cm (47 in)
• Width: 63.5 cm (25 in)
• Weight: Approx. 100 kg (220 lbs)
• Standard 2 Drawer: 6.8 kg (15 lbs)
• Standard Deep Drawer: 6.8 kg (15 lbs)
• Footprint: 119.4 x 63.5 cm (47 x 25 in)
• Floor Space: 7,582 sq cm (1,175 sq in)
• Mattress Size: 65 x 48 x 4 cm (25.5 x 19 x 1.5 in)
• Bed to Floor (fixed base): 83.8-107.9 cm (33-42.5 in) with 4 cm (1.6 in) fixed increments
• Bed to Floor (elevating): 83.8-107.9 cm (33-42.5 in)
• Bed Capacity: Patient weight up to 13.6 kg (30 lbs)
• Bed Tilting: ± 12° continuous tilt; Bubble level indicates level bed
Operational Environment Giraffe Warmer
• Operating Temp. Range: 18°C to 30°C
• Operating Humidity Range: 5% to 75%
• Storage Temp. Range: -25°C to +60°C
• Storage Humidity Range: 0% to 85%Resuscitation Unit
• Operating Temp. Range: 18°C to 41°C
• Operating Humidity Range: 0% to 90%
• Storage Temp. Range: -25°C to +60°C
• Storage Humidity Range: 0% to 95%
• Storage Pressure Range: 50 kPa to 106 kPa
Electrical Power Requirements • 115VAC, 50/60 Hz Models: 115V ± 10%; 6.6 Amps
• 220-240 VAC, 50/60 Hz Models: 230 V ± 10%; 3.3 Amps*
• Nominal Power Consumption: 520W at max. heater power, max. lighting
• Complies with IEC 60601-1 for electrical safety
Integrated Resuscitation Characteristics • Input pressure: 40-75 psi (275-517 kPa)
• Minimum Input Flow: 70 lpmVenturi Vacuum Subsystem
• Range: 0-150 mmHg
• Accuracy: +/- 5% of full scale

Adjustable PIP (For T-Piece Resuscitation System only)
• Maximum PIP: 45 ± 5 cm H2O
• Override: > 30 ± 4 cm H2O
• Flow Capacity: 15 lpm

Flowmeters
• Range: 0-15 lpm
• Accuracy:
0-5 lpm ± ½ lpm
5-10 lpm ± 1.5 lpm
10-15 lpm ± 2 lpm

Air/O2 Blender
• Range: 21-100% O2
• Accuracy: +/- 5% O2

Airway Pressure Manometer
• Display Range: -10 to 80 cm H2O
• Accuracy: +/- 5% of full scale

Integrated Scale Characteristics • Accuracy: +/- 10 g (0.35 oz)
• Range: 300 g to 8 kg (0.66 lbs to 17.6 lbs)
System Control Characteristics • Full Color Display: 16.5 cm (6.5 in)
• Observation Lights: 2 dimmable halogen lights (35 W each), 0-1500 lux average
• Procedure Light: 50 W, 2000 lux average; estimated life 3000 hrs based on manufacturer’s ratingWarmup Mode
• Automatically started at power up
• Heater operates at 100% power for 7 minutes, then reverts to 50% power
• Check patient alarms eliminated during Warmup mode

Microprocessor Based Control System
• Self-test function performed at power on and continuously during operation
• Heater Warmup Time at 100% Power: less than 3 minutes
• Heater Element: 360 watts
• Heater Output: adjustable from 0-100% in 5% increments
• Patient Control (Baby) Mode: 34°C to 37.5°C in increments of 0.1°C (95°F to 99.5°F)
• Manual Mode: Indicates manual mode heat selection range from 0 to 100% in 5% increments
• Temperature Measurement Accuracy: ± 0.3°C Resolution: ± 0.1°C
• Probe Accuracy: ± 0.1°C Range: 30-42°C (71.6-107.6°F)